19513727-V2.docx Print date: 13.01.14 Original Operating Manual High Temperature Circulator HT60-M2 HT60-M3 C.U.-cooling unit 1.951.3727-V2
Operator responsibility – Safety instructions 10
HT60-M2 / HT60-M3 11
Operator responsibility – Safety instructions 12 2.3. Warranty conditions JULABO GmbH warrants its products against defects in material or in workma
HT60-M2 / HT60-M3 13 2.4. Technical specifications HT60-M2 HT60-M2 & C.U. Working temperature range °C 70 ... 400 40 … 400 Temperature stabil
Operator responsibility – Safety instructions 14 HT60-M3 HT60-M3 & C.U. Working temperature range °C 70 ... 400 40 … 400 Temperature stabilit
HT60-M2 / HT60-M3 15 Electrical connections: Computer interface RS232 or RS485 or Profi Programmer input 0 - 10 V / 0 - 20 mA Temperatur recorder
Operator responsibility – Safety instructions 16 2.5. Cooling water connection Cooling water pressure (IN / OUT ) max. 4.5 bar Difference pressure
HT60-M2 / HT60-M3 17 3. Operating controls and functional elements M2 or M3 – control electronic 1-ESCP060180300420°C 1 Mains power switch, illu
Operating controls and functional elements 18 10 MULTI-DISPLAY (LED) temperature indication 11 Display of internal/external actual value /
HT60-M2 / HT60-M3 19 Rear view 1891616AMP16AMP16AMP16AMP16AMP16AMP4AMP4AMPHT60REG+E-PROG STAND-BY ALARMSERIALMADE IN GERMANY30 31 32 3335363738 30
Operating manual 2 Congratulations! You have made an excellent choice. JULABO thanks you for the trust you have placed in us. This operating manual ha
Operating controls and functional elements 20 HT60 High Temperature Circulator with C.U.-cooling unit (option) Front view: Cold zone Cooling
HT60-M2 / HT60-M3 21 40 Stand rod attachment with mouting screw; ∅12; M8 41 Connector for liquid level indicator or second expansion vessel; M
Safety notes for the user 22 4. Safety notes for the user 4.1. Explanation of safety notes In addition to the safety warnings listed above, warning
HT60-M2 / HT60-M3 23 • Employ suitable connecting tubing. • Avoid sharp bends in the tubing, and maintain a sufficient distance from surrounding wal
Safety notes for the user 24 Notice: Check the safety installations at least twice a year! • Excess temperature protection according to IEC 61010-2
HT60-M2 / HT60-M3 25 5. Preparations 5.1. Setup HT60CU -ESCPM2060180300420EXT°C High temperature circulator HT60 as well as High temperature ci
Preparations 26 5.2. Installation Caution: Installation of the C.U.-unit / Installation of the magnetic valve • Always turn off the unit
HT60-M2 / HT60-M3 27 Installation of the magnetic valve JULABO Order-No. 8 980 704 3 Magnetic valve • Fix the expansion tank (43) to the connect
Preparations 28 5.3. Bath fluids Caution: Carefully read the safety data sheet of the bath liquid used, particularly with regard to the fire point!
HT60-M2 / HT60-M3 29 5.4. Tubing Recommended tubing: Metal tubing, triple insulated, M16x1, Temperature range -100 °C ... +350 °C Order No. Length
HT60-M2 / HT60-M3 3 TABLE OF CONTENTS OPERATING MANUAL ...
Preparations 30 5.5. Power connection Caution: 1. Connect control cable (55) with control exit (35) of the control electronics M2 / M3 and latch
HT60-M2 / HT60-M3 31 4348495051555243a41 Filling 1. Attach a tube at the overflow nozzle (48) and lead it into a suitable vessel. 2. Remove the scre
Preparations 32 5.7. Degasifying During the filling process the pump fluidizes air bubbles into the tempering liquid. In the automatic degasifying m
HT60-M2 / HT60-M3 33 Setp.: 250.00°CIntAct: 250.00°CPower 50 %Control: intern 9. The DIALOG-DISPLAY (LCD) shows the standard indicat
Preparations 34 5.8. Countercooling Caution: Securely attach all tubing to prevent slipping. From a working temperature of 200°C on always work wit
HT60-M2 / HT60-M3 35 6. Operating procedures 6.1. Switching on / Selecting the language JULABO HT-30/60 Controller V 4.01-J Swi
Manual operation 36 7. Manual operation 7.1. Start - Stop Start: • Press the start/stop key . The MULTI-DISPLAY (LED) indi
HT60-M2 / HT60-M3 37 7.3. Warning functions )) )) )) More protection for your samples in the bath! An audible signal sounds in intervals when the a
Manual operation 38 7.4. Setting the safety temperature (with shutdown function) !+ A L A R M !ExtAct: xxx.xx°CTemp/level alarm ⇒
HT60-M2 / HT60-M3 39 7.5. Internal / external control EXT The High Temperature Circulator offers the possibility of internal temperature control i
Operating manual 4 8.7. Interface ...
Menu functions 40 8. Menu functions • Press the MENU key to enter the menu level. • Use the up/down cursor keys to select the desired submenu
HT60-M2 / HT60-M3 41 8.1. Configuration >Configuration Control param. Profile Start Int.Programmer Inputs/Outputs Limits Interface Temp.SensorC
Menu functions 42 • When the adjusted control parameters Xp, Tn and Tv are too high, this requirement may not be given with respect to on how much t
HT60-M2 / HT60-M3 43 Warning: For supervised or unsupervised operation with the AUTOSTART function, avoid any hazardous situation to persons or
Menu functions 44 8.2. Control parameters Configuration>Control param. Profile Start Int.Programmer Inputs/Outputs Limits Interface Temp.Sens
HT60-M2 / HT60-M3 45 JULABO HT-30/60 Controller V 4.01-J Note: The parameters >Xpu<, >CoSpeed< and >Dynamik< are only
Menu functions 46 Optimization instructions for the PID control parameters: °Ct The heat-up curve reveals inappropriate control settings. (example: w
HT60-M2 / HT60-M3 47 8.3. Start of a profile Configuration>Identif. off Setpoint keyb Autostart on 1. 2. The start m
Menu functions 48 Example: hour.min 6:00 h >hour.min 06:00 Day.Mon 15:11 Year 2000 Start yesStarttime: 06:00 Date: 15:11
HT60-M2 / HT60-M3 49 8.3.1. Interrupting a profile Setp. : xx.xx°CIntAct: xx.xx°CProf. : x Stp: x Remain: xx:xx:xx *** pause *** Interrupting
HT60-M2 / HT60-M3 5 Operating manual 1. Intended use The High Temperature Circulators HT60-M2 and HT60-M3 are to be employed especially for closed te
Menu functions 50 8.4. Integrated programmer The integrated programmer allows any desired temperature program sequences to be realized. Such a temp
HT60-M2 / HT60-M3 51 Int.Programmer>Edit Delete Print Set Clock>Edit Profile: 1 Step: 6 Setp: 100.00°C Time[h.m] 01:05 Examples: Profil
Menu functions 52 Int.Programmer Edit>Delete Print Set Clock>Delete Prof. 0 from Step 0 to Step 00 delete no Delete Prof.
HT60-M2 / HT60-M3 53 8.5. Analog inputs/outputs or This submenu enables setting of the input and output values for the progra
Menu functions 54 Examples: lowest temperature value: 10 °C highest temperature value 210 °C Fig. shows 200 °C scaled to paper width ris
HT60-M2 / HT60-M3 55 >EPROG: Current L Value 0.0°C H Value 300.0°CExtSet 50.0°CEPROG : 50.00°CIntAct: 50.00°CPower: 50 %Control:
Menu functions 56 8.6. Limits Inputs/Outputs>Limits Interface Temp.SensorLimits>IntMax 80.00°C IntMin 0.00°C Band 30K HeatingMax
HT60-M2 / HT60-M3 57 8.7. Interface Interface>Type RS232 Baudrate 4800 Parity no Handshake softw Inputs/Outputs Limits>Interface
Menu functions 58 Circulator (TT) ATC Int: ATC serves to compensate a temperature difference that might occur between circulator and a defined me
HT60-M2 / HT60-M3 59 9. Troubleshooting guide / Error messages !+ Whenever the microprocessor electronics registers a failure, a complete shutdown of
Operator responsibility – Safety instructions 6 2. Operator responsibility – Safety instructions The products of JULABO ensure safe operation when in
Troubleshooting guide / Error messages 60 • Safety sensor in the cooling zone: Cable of the safety sensor interrupted or short-circuited. • Safe
HT60-M2 / HT60-M3 61 10. Electrical connections Notice: Use shielded cables only. The shield of the connecting cable is electrically connected to th
Electrical connections 62 STAND-BY STAND-BY input (32) (for external emergency switch-off) Pin assignment: Pin Signal 1 not c
HT60-M2 / HT60-M3 63 EXT Socket for external Pt100 sensor (19) Pin assignment: Pin Signal 1 I+ 2 U+ 3 U- 4 I- The shield
Remote control 64 11.2. Communication with a PC or a superordinated data system Suitable terminal programs for communicating with a PC are: • MS-Wind
HT60-M2 / HT60-M3 65 11.3. List of commands When the RS485 interface is used, the instrument address stands in front of each command (Axxx_). in-comma
Remote control 66 in-commands: Asking for parameters or temperature values to be displayed. Command Parameter Response of circulator in_mode_03 non
HT60-M2 / HT60-M3 67 out commands: Setting parameters or temperature values. Command Parameter Response of circulator out_mode_02 0 No identificati
Remote control 68 Command Parameter Response of circulator out_par_14 xxx.xx Minimum internal temperature in case of cascade control Configuratio
HT60-M2 / HT60-M3 69 Error messages Description -12 TEMPERATURE MEASUREMENT ALARM Error in A/D converter. -13 WARNING : VALUE EXCEEDS TEMPERATURE LIMI
HT60-M2 / HT60-M3 7 Use The bath can be filled with flammable materials. Fire hazard! There might be chemical dangers depending on the bath medium us
Cleaning / repairing the unit 70 12. Cleaning / repairing the unit Caution: • Always turn off the unit and disconnect the mains cable from the powe
Operator responsibility – Safety instructions 8 2.2. EC Conformity
HT60-M2 / HT60-M3 9