19514866-V2.doc 26.07.13
English
Operating
manual
Cryo-Compact Circulators
The Economy-Series
CF30 CF40
JULABO GmbH
77960 Seelbach / Germany
Tel. +49 (0) 7823 / 51-0
Fax +49 (0) 7823 / 24 91
www.julabo.de
1.951.4866-V2 06/13
Vue de la page 0
1 2 3 4 5 6 ... 34 35

Résumé du contenu

Page 1 - Operating

19514866-V2.doc 26.07.13 English Operating manual Cryo-Compact Circulators The Economy-Series CF30 CF40 JULABO GmbH 77

Page 2

Operator responsibility – Safety instructions 10 Warning functions and safety installations Excess temperature protection adjustable from 0 °C ...

Page 3

Cryo-Compact Circulator 11 3. Safety notes for the user 3.1. Explanation of safety notes In addition to the safety warnings listed above, warnings

Page 4 - 1.1. Description

Safety notes for the user 12  Never operate the unit without bath fluid in the bath.  Pay attention to the thermal expansion of bath oil during he

Page 5

Cryo-Compact Circulator 13 Caution: The circulator may be used, for example, to control the temperature of fluids in a reactor. We do not know what s

Page 6 - 2.1. Disposal

Operating instructions 14 Operating instructions 4. Operating controls and functional elements Front view CF40 / CF30 Rear view CF40/ CF30 56a12

Page 7 - 2.3. Warranty conditions

Cryo-Compact Circulator 15 1 Mains power switch, illuminated 2 °C14010018022006020 Adjustable excess temperature protection according to IEC 61

Page 8

Preparations 16 5. Preparations 5.1. Installation  Place the unit on an even surface on a base made of nonflammable material.  Cooling ma

Page 9

Cryo-Compact Circulator 17 5.2. Temperature application to external, closed systems The Cryo-Compact Circulator is used for temperature application

Page 10

Preparations 18 5.3. Tubing Warning: Tubing: At high working temperatures, the tubing used for temperature control and for the cooling water suppl

Page 11 - 3.3. Safety instructions

Cryo-Compact Circulator 19 5.4. Bath fluids Caution: Carefully read the material safety data sheet of the bath fluid used, particularly with regard

Page 12

2 Congratulations! You have made an excellent choice. JULABO thanks you for the trust you have placed in us. This operating manual has been design

Page 13

Operating procedures 20 6. Operating procedures 6.1. Power connection Caution:  Only connect the unit to a power socket with an earthing contact (

Page 14 - Operating instructions

Cryo-Compact Circulator 21  Turn the mains switch (1) on (Switching on - see page 22)  Switch on unit. To do so press button for approx. 4 seco

Page 15

Operating procedures 22 6.3. Switching on / Start - Stop Switching on:  The Cryo-Compact Circulator is turned on and off with the mains swit

Page 16 - 5. Preparations

Cryo-Compact Circulator 23 6.5. Setting the temperatures  Setting can be carried out in the start/stop condition. 1. Press one of the keys fo

Page 17

Operating procedures 24 Warning: For supervised or unsupervised operation with the AUTOSTART function, avoid any hazardous situation to persons or

Page 18 - tubing 12 mm inner dia

Cryo-Compact Circulator 25 7. Safety installations Check the safety installations at least twice a year! (See page 13) 7.1. Excess temperature pro

Page 19 - 5.4. Bath fluids

Troubleshooting guide / Error messages 26 8. Troubleshooting guide / Error messages + Whenever the microprocessor electronics registers a failure,

Page 20 - 6. Operating procedures

Cryo-Compact Circulator 27  This message appear every 4 seconds. An acoustic signal sounds in regular intervals.  Compressor does not work. After

Page 21

Electrical connections 28 9. Electrical connections Notice: Use shielded cables only. The shield of the connecting cable is electrically connected t

Page 22

Cryo-Compact Circulator 29 10. Remote control 10.1. Setup for remote control 9651 SERIAL 1. Check the interface parameters for both interfaces (o

Page 23 -  AUTOSTART on

Cryo-Compact Circulator 3 TABLE OF CONTENTS Operating manual ...

Page 24

Remote control 30 Important times for a command transmission:  To ensure a safe data transfer, the time gap between two commands should be at least

Page 25 - 7. Safety installations

Cryo-Compact Circulator 31 in commands: Asking for parameters or temperature values to be displayed. Command Parameter Response of recirculating c

Page 26

Remote control 32 Error messages Description -09 COMMAND NOT ALLOWED IN CURRENT OPERATING MODE Invalid command in current operating mode. -10 VALUE

Page 27

Cryo-Compact Circulator 33 11. JULABO Service – Online remote diagnosis JULABO Cryo-compact circulator s of the HighTech series are equipped with

Page 28 - 9. Electrical connections

Draining 34 12. Draining Notice: Do not drain the bath fluid while it is hot! Check the temperature of the bath fluid prior to draining (by switc

Page 29 - 10. Remote control

Cryo-Compact Circulator 35 T16AT16ASERIALOFFAUST16AT16AC To maintain the full cooling performance, clean the condenser (C) from time to time. 1. S

Page 30 - 10.3. List of commands

Operating manual 4 Operating manual 1. Intended use JULABO Cryo-Compact Circulators have been designed for temperature application to specific fluids

Page 31

Cryo-Compact Circulator 5 If you have any questions concerning the operation of your unit or the information in this manual, please contact us! Conta

Page 32

Operator responsibility – Safety instructions 6 When using hazardous materials or materials that could become hazardous, the operator must affix the e

Page 33

Cryo-Compact Circulator 7 2.2. EC Conformity The products described in the operating instructions conform to the requirements of the following Europe

Page 34 - 12. Draining

Operator responsibility – Safety instructions 8 2.4. Technical specifications CF30 Working temperature range °C -30 ... +150 Temperature stabil

Page 35

Cryo-Compact Circulator 9 CF40 Working temperature range °C -40 ... +150 Temperature stability °C ±0.03 Temperature selection digital vi