19514871-V2.doc 26.07.13
Operating Manual
Cryo-Compact Circulators
The HighTech Series
CF31
CF41
1.951.4871-V2 06/13
JULABO GmbH
77960 Seelbach / Germany
Tel. +49 (0) 7823 / 51-0
Fax +49 (0) 7823 / 24 91
www.julabo.de
Vue de la page 0
1 2 3 4 5 6 ... 75 76

Résumé du contenu

Page 1 - Operating Manual

19514871-V2.doc 26.07.13 Operating Manual Cryo-Compact Circulators The HighTech Series CF31 CF41 1.951.4871-V2 06/13 JULAB

Page 2

Operator responsibility – Safety instructions 10 CF41 Working temperature range °C -40 ... +200 Temperature stability °C ±0.02 Temperature

Page 3

Cryo-Compact Circulators 11 Electrical connections: Computer interface RS232 / RS485 External sensor Pt100 Optional: (Order No. 8900100 El

Page 4

Safety notes for the user 12 3. Safety notes for the user 3.1. Explanation of safety notes In addition to the safety warnings listed above, warnin

Page 5 - 1. Intended use

Cryo-Compact Circulators 13 overflowing of the fluid.  Prevent water from entering the hot bath oil.  Do not drain the bath fluid while it is hot!

Page 6

Safety notes for the user 14 The following questions should help to recognize possible dangers and to reduce the risks to a minimum.  Are all tube

Page 7 - 2.1. Disposal

Cryo-Compact Circulators 15 Operating instructions 4. Operating controls and functional elements Front view CF41 / CF31 Rear view CF41/ CF31 12

Page 8 - 2.3. Warranty conditions

Operating controls and functional elements 16 1 Mains power switch, illuminated 2 °C14010018022006020 Adjustable excess temperature protection acco

Page 9

Cryo-Compact Circulators 17 4.3 Display for the adjusted pump pressure stage in the -OFF- mode. Display for the effective pump pressure stage (rota

Page 10

Operating controls and functional elements 18 Option: Electronic module Order No. 8 900 100 The Cryo-Compact Circulator automatically recognize

Page 11

Cryo-Compact Circulators 19 5. Preparations 5.1. Installation  Place the unit on an even surface on a base made of nonflammable material.  Cool

Page 12 - 3.3. Safety instructions

2 Congratulations! You have made an excellent choice. JULABO thanks you for the trust you have placed in us. This operating manual has been design

Page 13

Preparations 20 Return flow safety device For this reason, shut-off valves can be integrated in the loop circuit. Order No. Description 8 970 456

Page 14

Cryo-Compact Circulators 21 5.3. Tubing Warning: Tubing: At high working temperatures, the tubing used for temperature control and for the coolin

Page 15 - Operating instructions

Preparations 22 5.4. Bath fluids Caution: Carefully read the material safety data sheet of the bath fluid used, particularly with regard to the fi

Page 16 - Remote control

Cryo-Compact Circulators 23 6. Operating procedures 6.1. Power connection Caution:  Only connect the unit to a power socket with an earthing conta

Page 17

Operating procedures 24 Important:  When using a bath fluid, the change in volume in case of change in temperature has to be respected. Fill in a li

Page 18 - REG+E-PROG

Cryo-Compact Circulators 25 7. Manual operation 7.1. Start - Stop 24.60S 150.00E ----- Setpoint1: 150.00°CIntAct : 24.60°CPower : 80 %

Page 19 - 5. Preparations

Manual operation 26 7.3. Setting the temperatures Factory setting: Setpoints>Setpoint1: 20.00°C Setpoint2: 37.00°C Sezpoint3: 70.00°C Press

Page 20

Cryo-Compact Circulators 27 7.4. Safety installations, warning functions Check the safety installations at least twice a year! (See page 14) 7.4.1

Page 21 - 5.3. Tubing

Manual operation 28 Recommendation: Set the excess temperature protection >SafeTemp< at 5 to 10 °C above the working temperature setpoint. W

Page 22 - 5.4. Bath fluids

Cryo-Compact Circulators 29 OverTemp 87.0°CSubTemp 83.0°CSetpoint3 85.0°C  The warning functions are only triggered when the actual bath temperatur

Page 23 - 6. Operating procedures

Cryo-Compact Circulators 3 TABLE OF CONTENTS Operating manual ...

Page 24 - * Version 1.02-41 *

Manual operation 30 7.4.2. Early warning system, low level protection (patented) This low level protection is independent of the control circuit a

Page 25 - 7. Manual operation

Cryo-Compact Circulators 31 8. Menu functions The term „menu functions“ refers to adjustments such as  Configuration of the Cryo-Compact Circu

Page 26

Menu functions 32 8.1. Configuration >Configuration Control Profile StartMain menuConfiguration>Setpoint : keypad /RS232/ext.Pt100/eprog Au

Page 27 - 1. Press the key

Cryo-Compact Circulators 33 eprog : 50.0°CIntAct : 24.7°CPower : 80 %Control : intern eprog – Can only be adjusted when an elec

Page 28

Menu functions 34 OFF-Mode Normally the circulating pump is switched via the start/stop signal. However, if circulation should be maintained also for

Page 29 - Setpoint3 85.0°C

Cryo-Compact Circulators 35 Warning: The working temperature range of the Cryo-Compact Circulator is determined during configuration. If set to

Page 30 - Replenish the bath

Menu functions 36 8.2. Control The Cryo-Compact Circulator is conceived for internal and external temperature control. Switching is carried out in

Page 31 - >Configuration

Cryo-Compact Circulators 37  Sensor calibration of the external Pt100 sensor is carried out in the menu >ATCalibration<, with the > Status&l

Page 32 - R“ illuminates

Menu functions 38 Dynamic: This parameter affects the temperature pattern only in case of internal control. °Ctaperiodicnormalsetpointtemp. stability

Page 33 - Control : intern

Cryo-Compact Circulators 39 Setting range: internal / external 3 ...9999 s Resetting time >Tn< (Integral component) Compensation of the rema

Page 34

4 8. Menu functions ... 31 8.1. Configura

Page 35

Menu functions 40 8.3. Start of a profile The start menu of the integrated programmer allows calling up and defined starting of one of six previous

Page 36 - 8.2. Control

Cryo-Compact Circulators 41 Examples: -OFF-S 25.00E ----- Setpoint3: 80.00°CIntAct : 80.00°CPower : 2%Control : intern Se

Page 37

Menu functions 42 Examples: Setp. : 125.8°CIntAct: 124.7°CProf. : 4 Step: 12Remain: 00:03:45 Setp. : 125.8°CIntAct: 124.

Page 38

Cryo-Compact Circulators 43 >Edit Profile: 1**Step is deleted ** Setp: °C Time [hh:mm]Edit >Edit Profile: 1** Step i

Page 39

Menu functions 44 8.3.2. Interrupting a profile Setp. : 125.80°CIntAct: 124.70°CProf. : 4 Step:12 *** pause ***  Press the start/stop

Page 40 - 8.3. Start of a profile

Cryo-Compact Circulators 45 8.3.4. Termination of a profile Setp : 125.80°CIntAct: 124.70°CProf. : 4 Step:12Remain: 00:03:45  >

Page 41 - *** wait ***

Menu functions 46 8.4. Integrated programmer The integrated programmer allows any desired temperature program sequences to be realized. Such a temp

Page 42 - All: hours:minutes:seconds)

Cryo-Compact Circulators 47 Int. Programmer>Edit Delete Print Set Clock Edit Compile profiles Display sections Important: Also a running prof

Page 43 - Time [hh:mm] 01:05

Menu functions 48 >Edit Profile: 1 Step: 3 Setp: 50.00°C Time [hh:mm] 00:02Edit Display sections (Step): 1 Use the keys

Page 44

Cryo-Compact Circulators 49 8.5. Analog inputs/outputs ALARMREG+E-PROGSTAND-BY  In order to use the analog inputs and outputs, the Cryo-Compact Cir

Page 45

Cryo-Compact Circulators 5 Operating manual 1. Intended use JULABO Cryo-Compact Circulators have been designed for temperature application to specifi

Page 46 - 8.4. Integrated programmer

Menu functions 50 Menu level 2 Inputs/Outputs>Chan.1 : ActInt EPROG : Setpoint Ext.StBy : inactiv /activ AlarmOut : StdBy Chan.2 :

Page 47 - Set Clock

Cryo-Compact Circulators 51 8.5.1. REG+E-PROG – temperature recorder outputs CHANNEL 1CHANNEL 3CHANNEL 2REG+E-PROGEPROG REG+E-PROG (11): Outputs o

Page 48 -  Lines 1 to 3:

Menu functions 52 8.5.2. REG+E-PROG - programmer input Menu >Configuration< Configuration>Setpoint : eprog Keypad/RS232/extPt100 Autost

Page 49 - 8.5. Analog inputs/outputs

Cryo-Compact Circulators 53 Eprog : 50.0°CIntAct : 24.7°CPower : 80 %Control : intern  After returning the LCD display to stan

Page 50

Menu functions 54 8.5.3. STAND-BY input Ext. StBy (10) (for external emergency switch-off) (Connector see page 67) Possible parameters: inactiv -

Page 51 - 10V = 210°C

Cryo-Compact Circulators 55 8.5.4. ALARM output ALARM Alarm output (9) (for external alarm signal) This socket is a potential-free change-ove

Page 52 - (Liters/Minute)

Menu functions 56 8.6. Limits The menu >Limits< allows for setting the minimum and maximum values for all important setting ranges and capacit

Page 53 - Eprog : 50.0°C

Cryo-Compact Circulators 57 >SafeTemp:< IntMax: and IntMin: Limit setting Setting range: see working temperature range page 9

Page 54

Menu functions 58 8.7. Serial interface For communication between Cryo-Compact Circulator and a PC or a superordinated process system the interface

Page 55 - Menu >Inputs/Outputs<

Cryo-Compact Circulators 59 8.8. ATC Absolute Temperature Calibration, 3-point calibration ATC serves to compensate a temperature difference that m

Page 56 - 8.6. Limits

Operator responsibility – Safety instructions 6 2. Operator responsibility – Safety instructions The products of JULABO ensure safe operation when i

Page 57

Menu functions 60 Type: A >1-point<, >2-point < or >3-point < calibration can be carried out. First geometrically define the locat

Page 58 - 8.7. Serial interface

Cryo-Compact Circulators 61 1. Use the keys to select the menu item. 2. With the keys select the parameter (parameter blinks) or use the n

Page 59

Menu functions 62 8.9. Setting the pump pressure Examples: Adjusted Effective The pressure of the circulating pump is adjustable in four

Page 60

Cryo-Compact Circulators 63 9. Troubleshooting guide / Error messages XXXXXALARMCODE 01 + Alarm with a complete shutdown of the unit Whenever the m

Page 61 - Setpoint3: 70.00°C

Troubleshooting guide / Error messages 64 ALARM CODE 06  Difference betweenworking- and safety-temp. Bigger thanpermissible value.  Defect of the

Page 62 - Pump level: 1 /2/3/4

Cryo-Compact Circulators 65 ALARM CODE 38  Ext. Sensor inputwithout signal butsetpoint programmingset to ext. Pt100.  Ext. Pt100 sensor input with

Page 63

Electrical connections 66 10. Electrical connections Notice: Use shielded cables only. The shield of the connecting cable is electrically connected

Page 64 - The internal I2C

Cryo-Compact Circulators 67 4132 ext. Pt100 Socket for external Pt100 sensor 1423Shield PlugLook on soldering side.Pt100 Pin assignment: Pin S

Page 65 - Ext. Sensor input

Remote control 68 11. Remote control 11.1. Setup for remote control 9651 SERIAL -OFF-S xxxxI xxxxR 1. Check the interface parameters for both int

Page 66 - 10. Electrical connections

Cryo-Compact Circulators 69 The commands are divided into in or out commands. in commands: asking for parameters to be displayed out commands: sett

Page 67

Cryo-Compact Circulators 7 sheets). Insufficient ventilation may result in the formation of explosive mixtures. Only use the unit in well ventilated a

Page 68 - 11. Remote control

Remote control 70 Command Parameter Response of Cryo-Compact Circulator out_mode_03 0 Set external programmer input to voltage. Voltage 0 V ..

Page 69 - 11.3. List of commands

Cryo-Compact Circulators 71 in commands: Asking for parameters or temperature values to be displayed. Command Parameter Response of Cryo-Compact C

Page 70

Remote control 72 Command Parameter Response of Cryo-Compact Circulator in_par_11 none Tn control parameter of the cascade controller. in_par_12

Page 71

Cryo-Compact Circulators 73 11.5. Error messages Error messages Description -01 LOW LEVEL ALARM Low liquid level alarm -03 EXCESS TEMPERATURE WARNING

Page 72 - 11.4. Status messages

JULABO Service – Online remote diagnosis 74 12. JULABO Service – Online remote diagnosis JULABO Cryo-Compact Circulator s of the HighTech series a

Page 73 - 11.5. Error messages

Cryo-Compact Circulators 75 13. Installation of electronic module Caution: Always turn off the unit and disconnect the mains cable from the power

Page 74

Cleaning / repairing the unit 76 14. Cleaning / repairing the unit Caution: Always turn off the unit and disconnect the mains cable from the power

Page 75

Operator responsibility – Safety instructions 8 2.2. EC Conformity The products described in the operating instructions conform to the requirements

Page 76

Cryo-Compact Circulators 9 2.4. Technical specifications CF31 Working temperature range °C -30 ... +200 Temperature stability °C ±0.02 Temper