21
Selecting the working temperature:
Press the setpoint key
and then enter .
Press the setpoint key
and then enter .
7.3. Warning functions
  
More protection for your samples in the bath!
An audible signal sounds in intervals when the actual
temperature value exceeds one of the set limits (patented).
Setting the high temperature limit:
Press the key .
The value previously set appeears on the DIALOG-
DISPLAY (LCD) (example: 305.00°C). A flashing segment
indicates that a value needs to be entered.
Use the keypad to enter the new value (example: 39.00
°C).
Press enter to store the value.
Setting the low temperature limit:
Press the key .
Follow the instructions
for (example: 35.00 °C).
Note:
The warning functions will only be triggered when the actual bath temperature,
after start from the „OFF“ or „rOFF“ mode, lies within the set limits for 3 seconds.