19514802-V2.doc 26.07.13 English Operating manual Recirculating coolers FC600 FC600S FCW600 FCW600S FC1200 FC1200S FCW1200 FCW1200S FC
Operator responsibility – Safety instructions 10 Safety installations (adjustable via LED): High temperature protection °C -25 ... 85 Low temperat
Recirculating coolers 11 2.5. Cooling water connection Only for water cooled models - FCW: Cooling water pressure (IN / OUT ) max. 6 bar Differen
Safety notes for the user 12 3. Safety notes for the user 3.1. Explanation of safety notes In addition to the safety warnings listed above, warnin
Recirculating coolers 13 Make sure you read and understand all instructions and safety precautions listed in this manual before installing or opera
Operating controls and functional elements 14 4. Operating controls and functional elements 1. Mains power switch, illuminated 7. Filling funne
Recirculating coolers 15 4. Keys for actual values (LED 1) A Key - Indication of feed temperature B Key - Indication of return temperature
Operating controls and functional elements 16 Rear 12.1 Pump connector: Feed 12.2 Pump connector: Return 13. Overflow port for bath tank 14. Con
Recirculating coolers 17 5. Operation 5.1. Preparations 12.1 12.2 M16x1 ! 13. 14. 16.2 12.1 16.1 12.2 16.1 16.2 ¾&
Operation 18 Notice: Cooling water circuit Risk of oil leaking from the cooling circuit (compressor) of the recirculating cooler into the cooling w
Recirculating coolers 19 5.3. Bath fluids Caution: No liability for use of other bath liquids! Please contact JULABO before using other than recom
2 Congratulations! You have made an excellent choice. JULABO thanks you for the trust you have placed in us. This operating manual has been designed
Operation 20 5.5. Power connection Caution: Only connect the unit to a power socket with earthing contact (PE – protective earth)! We disclaim a
Recirculating coolers 21 5.7. Draining Notice: Do not drain the bath fluid while it is hot or cold! Check the temperature of the bath fluid pr
Manual operation 22 6. Manual operation 6.1. Switching on ACTUAL SET Turn on the mains power switch (1.). An illumi
Recirculating coolers 23 6.2.1. Automatic / non-automatic start mode NOTE: The recirculating cooler has been configured and supplied by JULABO accord
Manual operation 24 6.3. Setting the setpoint temperatures Set the setpoints before or after starting the unit. Press the setpoint keys (E, F, G,
Recirculating coolers 25 6.3.2. Setting the control ratio for feed/return flow temperature 50 (%)21°C 22°C20°C-0.5°C +0.5°C5050 0 .. 100 (%)21°C22.5°
Manual operation 26 6.3.3. Setting the safety temperatures F G This safety function is independent of the control circuit. Press the d
Recirculating coolers 27 6.4. PID control parameters For internal and external control two separate parameter sets are available. The PID control
Manual operation 28 Optimization instructions for the PID control parameters: The heat-up curve reveals inappropriate control settings optimum s
Recirculating coolers 29 7. Trouble shooting guide + ACTUAL SET Whenever the microprocessor electronics registers a failure, an alarm i
Recirculating coolers 3 Table of Contents 1. Intended use ...
Electrical connections 30 7.1. Other error messages ACTUAL SET Incorrect/invalid entry. Value too small or too large, or function not
Recirculating coolers 31 Serial interface (K) 9651 This interface is a 9-pole connector: Pin 2 RxD Receive Data Pin 3 TxD Transmit Data Pin 5 G
Remote control 32 9. Remote control 9.1. Communication with a PC or data system ACTUAL SET For remote control, under the menu item OP (Op
Recirculating coolers 33 9.2. List of commands Command Parameter Response of recirculating cooler version no Number of software version status
Remote control 34 9.3. Status messages Message Description - Recirculating cooler ... 00 MANUAL STOP ... in condition "OFF" (LOCAL) 01 MA
Recirculating coolers 35 10. Menu functions Set the parameters for the recirculating cooler via the configuration or calibration level. 10.1. Sel
Menu functions 36 10.3. Adjustable parameters ACTUAL SET Set the parameters for the following menu items in the configuration level: O
Recirculating coolers 37 11. Cleaning / repairing the unit Caution: Always turn off the unit and disconnect the mains cable from the power source b
Intended use 4 1. Intended use JULABO recirculating coolers have been designed for temperature application to specific fluids. The pump connections c
Recirculating coolers 5 The operator is responsible for the qualification of the personnel operating the units. The personnel operating the units
Operator responsibility – Safety instructions 6 Use For the use according to the intended purpose, special material requirements have to be respected
Recirculating coolers 7 2.2. EC Conformity The products described in the operating instructions conform to the requirements of the following Europea
Operator responsibility – Safety instructions 8 2.4. Technical specifications FC(W)600 FC(W)600S Working temperature range °C -20 ... 80 -10
Recirculating coolers 9 FCW1600 FCW1600S Working temperature range °C -20 ... 80 -15 ... 80 Cooling capacity °C 20 5 -10 20